Sensor specifications | |
---|
Measurement range (trace range - liquid) | 0.1 - 250 µg/l 0.1 - 250ppb |
Measurement range (trace range - gas) | 3ppm - 6000ppm |
Resolution (trace range - liquid) | 0.1 µg/l 0.1ppb |
Resolution (trace range - gas) | 3 ppm |
Detection limit (trace range - liquid) | 0.1 µg/l 0.1ppb |
Detection limit (trace range - gas) | 3 ppm |
Measurement range (standard - liquid) | 0.01-20 mg/l 0.1-200% Sat |
Measurement range (standard - gas) | 1 mbar - 400 mbar |
Resolution (standard - liquid) | 0.01 mg/l 0.1% Sat
|
Resolution (standard - gas) | 1 mbar |
Detection limit (standard - liquid) | 0.01 mg/l 0.1% Sat
|
Detection limit (standard - gas) | 1 mbar |
Accuracy | +/- 1% upper range value |
Response time (t90) | < 60 s (air to N2 @ 25°C) |
Operating conditions | |
---|
Opration temperature (probe tip) | 0 to +135 °C |
Storage temperature | -20 to +60 °C |
Protection class | waterproof according to IP67 / NEMA6 |
Design characteristics | |
---|
Temperature compensation | automatically by integrated temperature sensor 0 - 80 °C < 2 % |
Pressure compensation | automatically via input 500 - 1100 hPa |
Salinity correction | automatically via input 0,01 - 70 SAL |
Sensor parts touched by media | stainless steel, silicon (FDA), sapphire |
Seals touched by media | silicon (FDA) |
Probe diameter | 12 mm |
Probe length | 120 mm, 160mm, 225mm, 325mm, 425mm customized length on request |
Connection | PG 13.5 |
Interfaces | RS485 (Modbus RTU), 2x 0-20mA or 4-20mA, with watchdog >24mA, 2x digital I/Os |
power | 12-36VDC @ 2VA max. |